8.2.3. Initial Pressurization Value of S-BPM
An initial pressurization value of S-BPM can be set beforehand.
160, 180, 210, 240, 270, AUTO [mmHg]
If AUTO is specified, a proper initial pressurization value is
selected automatically. The factory setting is 180 mmHg.
Step 1. Set the AUTO switch to "OFF".
Step 2. If the display is hidden, press the START/STOP
or EVENT switch to return to the waiting mode display.
Step 3. Select a pressurization value using the START/STOP switch.
8.3. A-BPM Preset Programs
The initial settings can be configured using the following methods.
Using the switches on the recorder.
Using a dedicated peripheral that is connected to the recorder
using the USB cable.
A-BPM can only be used while the automated measurement can be
performed.
Specify the start time and interval calculated from the time
that the AUTO switch is set to "ON".
It is necessary to specify them again when using another A-BPM.
S-BPM program
Pressurization value
Current time
Note
50
AUTO switch to "OFF"