Blood Pump (Single Needle), Optional - Fresenius Medical Care 4008 E Technical Manual

Hemodialysis machine
Table of Contents

Advertisement

1.2.2

Blood pump (single needle), optional

Essentially, the blood pump (single needle) is identical with the arterial blood pump. The
difference lies in the single needle control. During SN operation, the pressure outlet of the
compliance vessel is connected to the pressure connector of the SN pump. The pressure
transducer is protected by a hydrophobic filter both in the external and the internal tubing system.
The SN stroke volume can be set within a range from 10 ml to 50 ml in increments of 5 ml.
The lower changeover point is fixed to 75 mmHg.
The upper changeover point depends on the stroke volume:
Stroke volume (ml)
Changeover point (mmHg)
± 7 mmHg
10
15
20
25
110
130
150
172
Fresenius Medical Care 4008 4/08.03 (TM)
30
35
40
45
195
219
244
270
50
299
1-79

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4008 b4008 h4008 s

Table of Contents