Error description:
Error message
F01 Arterial
F02 Arterial
F95 Arterial
Description
With detuning in negative direction, the change achieved on the arterial
display is less than 100 mmHg (2 sec).
– Acknowledgement (P_ART, X348A/7) → X633L/B12, insufficient
voltage change.
– Test detuning defective (PA_DET, X632/A17) → X348A/9.
With detuning in positive direction, the change achieved on the arterial
display is less than 100 mmHg (2 sec).
– Acknowledgement (P_ART, X348A/7) → X633L/B12, insufficient
voltage change.
– Test detuning defective (PA_DET, X632/A17) → X348A/9.
System error.