Fresenius Medical Care 4008 E Technical Manual page 224

Hemodialysis machine
Table of Contents

Advertisement

O: CPU1_CLR_ALARM
Clearing of the alarm
IC 11/18
Check: generate a heparin pump
alarm; by setting CLR_ALARM from 0
to 1, the alarm is cleared.
O: CPU1_HOT_RINSE
Changeover, hot rinsing
IC 11/17
O: CPU1_TEST_BATT
Battery test
IC 11/16
O: CPU1_CPU_AKKU
Battery relay
IC 11/15
O: CPU1_HEAT_OFF
Heater blocking
IC 11/14
O: CPU1_STAFF_CALL
Staff call
IC 11/13
O: CPU1_TL_RED
Traffic light red
IC 11/12
(The pertinent status indicator on the
front panel lights simultaneously.)
O: CPU1_BPST_ART
Special control of the arterial blood
pump
IC 6/19
O: CPU1_BPSST_ART
Arterial blood pump system stop
IC 6/18
O: CPU1_CLAMP_CTRL
LD clamp control
IC 6/17
Preparation: LD in no alarm state
O: CPU1_BPST_VEN
Special control of the venous blood
pump
IC 6/16
O: CPU1_BPSST_VEN
Venous blood pump system stop
IC 6/15
Preparation:
Set SNST (CPU1: WR DIGIT. OUTP)
to 1
Check: apply pressure to the SN blood
pump; with BPSST_VEN, the blood
pump can be switched on/off.
O: CPU1_BL_ALARM
Changeover, blood pump rate
Single needle/HDF
IC 6/14
O: CPU1_TL_YELLOW
Traffic light yellow
IC 6/13
(The pertinent status indicator on the
front panel lights simultaneously.)
O: CPU1_TL_GREEN
Traffic Light green
IC 6/12
The pertinent status indicator on the
front panel lights simultaneously.
Fresenius Medical Care 4008 4/08.03 (TM)
5-25