Fresenius Medical Care 4008 E Technical Manual page 227

Hemodialysis machine
Table of Contents

Advertisement

O: CPU2_4066_ENABLE_4
Analog switch for PA_DET
IC 20/12 (X632/A17)
O: CPU2_4066_ENABLE_5
Analog switch, not used
IC 27/13
O: CPU2_4066_ENABLE_6
Analog switch for COND_DET
IC 20/5 (X632/A21)
O: CPU2_4066_ENABLE_7
Analog switch for BLL_DIM
IC 27/6
O: CPU2_4066_ENABLE_8
Analog switch for BLL
IC 27/12
O: CPU2_SN_ART
Single needle control, arterial
X632/A15
O: CPU2_LDSA
Attenuation of LD ultrasonic sensor
X632/C16
Preparation:
LD alarm-free and set CLAMP_CTRL
(CPU1: WR DIGIT. OUTP) to 1.
Check: by setting LDSA to 1, the
clamp at the LD closes.
O: CPU2_ODSA
Attenuation of LD optical sensor
X632/C15
5-28
Fresenius Medical Care 4008 4/08.03 (TM)
O: CPU2_CLAMP_CTRL
Clamp control, air detector
X632/C10
Preparation: LD in no alarm state
O: CPU2_NC5
Not used
X632/B25
O: CPU2_NC7
Not used
X632/B10
O: CPU2_BLL_DET
Detuning of blood leak detector
X632/A25
O: CPU2_SN_EN
Release of single needle
X632/C19
O: CPU2_NC10
Not used
X632/B4
O: CPU2_V26
Valve 26
X632/A6
O: CPU2_V42
Not used
X632/C4
O: CPU2_V43
Valve 43
X632/C5