Fresenius Medical Care 4008 E Technical Manual page 30

Hemodialysis machine
Table of Contents

Advertisement

Error description:
Error message
F01 Display
F02 Display
F95 Display
Description
CPU1 failed to start the display test within 5 sec.
– The "test started" information transmitted via the serial interface is
missing.
CPU1 failed to complete the display test within 120 sec.
– The "test completed" information transmitted via the serial interface
is missing.
System error.