Part 12.4: Determine the concentrate pump volume
(
Press ALARMTONE MUTE
4-22
Fresenius Medical Care 4008 4/08.03 (TM)
CONC. PUMP VOL det.
Confirm
key
START MAND. FILLING?
Override
(Tone Mute)
key
MAND. FILLING active
Time left displayed
in the field UF volume
after 9 seconds
conc pulses = 100
Override
(Tone Mute)
key
ACKNOWLEDGED
approx. 3 sec
)
Press OVERRIDE
Override
(Tone Mute)
key
conc puls left = 100
The concentrate pump
runs for 100 strokes.
The display counts down.
Then determine and record
the volume removed
by the pump.
Notes:
100 strokes are factory-set. This setting can be
changed by pressing the
(depending on the graduated cylinder used).
However, when returning to "CAL. MIXING-SYSTEM",
the display will indicate the factory setting again.
Check the volume and, if necessary,
repeat the procedure.
Note
Accuracy of the measuring cylinder: ±0.5 %.