Lp 632 (Cpu 2) Dip Switch Array - Fresenius Medical Care 4008 E Technical Manual

Hemodialysis machine
Table of Contents

Advertisement

3.1.3
P.C.B. LP 632 (CPU 2) DIP switch array 2
LP632
SW1
Switch
Switch 1
Switch 2
Switch 3
Switch 4
Switch 5
Switch 6
Switch 7
Switch 8
The basic position upon delivery is shown in italics.
For "not used" the switch must be set to OFF.
3-14
Fresenius Medical Care 4008 4/08.03 (TM)
Position
Function
ON
DIASAFE / DIASAFE
OFF
DIASAFE / DIASAFE
ON
ON-LINE-HDF on
OFF
ON-LINE-HDF off
ONLINE plus ™ on
ON
OFF
ONLINE plus ™ off
ON
Advanced hydraulics
OFF
Basic hydraulics
ON
V39 Test off
OFF
V39 Test on
ON
Fast heating HDIS deactivated
OFF
Fast heating HDIS activated
ON
Not used
OFF
ON
Not used
OFF
SW2
1
2
3
4
5
6
ON
OFF
®
plus on
®
plus off
7
8