8.21
P.C.B. LP 645 Position sensor membrane pump
8.21.1
Description
General information
P.C.B. LP 645 comprises the circuit for the recognition of the end position of the membrane pump.
Description
IC 1 and IC 2 are provided as two optical sensors. They are configurated such that their outputs
are linked by an AND operation. Their joint output is X5.
X5 turns to L level when the membrane of the pump is located in the front dead point (initialization
point).
Fresenius Medical Care 4008 4/09.03 (TM)