6.1
Overview Setup menu settings
Menu item
SET ALARM/WARN TIME
Menu item
SETUP CLEANING PGM
CLEANING Times
Continued on the next page
Submenu
Set ART-AL DELAYTIME
Set VEN-AL DELAYTIME
Set FLOW-OFF W-TIME
Set UF-WARNING-TIME
Set MUTE-TIME
Submenu
Rinsing TIME
Hotrinse TIME
Disinfection TIME
Rinsing Free TIME
Hot-Disinf TIME
Mandatory Rinse TIME
CITRO-Mandat-Ri-Time
INTEGRATED-HR Time
Fresenius Medical Care 4008 4/08.03 (TM)
Default value
Value range
5 sec
0 – 5 sec
5 sec
0 – 5 sec
30 min
30 – 60 min
10 min
10/30 min
1 min
1 – 2 min
Default value
Value range
15 min
5 – 30 min
15 min
15 – 30 min
10 min
10 – 20 min
CPU1: DIP switch array 1, SW8 set to OFF
(Test flow 800 ml/min)
3 min
3 – 10 min
for CDS:
5 min
5 – 10 min
CPU1: DIP switch array 1, SW8 set to ON
(Test flow 500 ml/min)
4 min
4 – 10 min
for CDS:
6 min
6 – 10 min
10 min
10 – 20 min
CPU1: DIP switch array 1, SW8 set to OFF
(Test flow 800 ml/min)
15 min
15 – 30 min
for ON-LINE-HDF:
20 min
20 – 30 min
for ONLINE plus ™:
17 min
17 – 30 min
CPU1: DIP switch array 1, SW8 set to ON
(Test flow 500 ml/min)
15 min
15 – 30 min
for ON-LINE-HDF:
20 min
20 – 30 min
for ONLINE plus ™:
20 min
20 – 30 min
CPU1: DIP switch array 1, SW8 set to OFF
(Test flow 800 ml/min)
10 min
10 – 25 min
for ON-LINE-HDF:
20 min
20 – 25 min
for ONLINE plus ™:
17 min
17 – 25 min
CPU1: DIP switch array 1, SW8 set to ON
(Test flow 500 ml/min)
10 min
10 – 25 min
for ON-LINE-HDF:
20 min
20 – 25 min
for ONLINE plus ™:
20 min
20 – 25 min
15 min
15 – 40 min
Resolution
1 sec
1 sec
15 min
20 min
1 min
Resolution
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min
1 min