Flow Pump Pressure - Fresenius Medical Care 4008 E Technical Manual

Hemodialysis machine
Table of Contents

Advertisement

3.3.4

Flow pump pressure

Measuring equipment:
Place of measurement:
Prerequisite:
Check/adjustment:
– Check the pressure of the flow pump
Connect the measuring equipment to the measurement port C.
Turn the water supply off; water alarm; balancing chamber inactive.
Measure the pressure of the flow pump.
Rated value of the flow pump pressure: 2.0 to 2.1 bar
If it deviates from the rated value, the pressure of the flow pump must be adjusted.
– Adjust the pressure of the flow pump
Use the relief valve (78) to adjust the rated value.
Measurement port C
Relief
valve 78
4008 B/S
UMED, HMED or pressure gauge
Hydraulics, measurement port C
A dialysate flow of 800 ml/min must have been preselected.
Fresenius Medical Care 4008 4/09.03 (TM)
Fresenius Medical Care 4008 4/08.03 (TM)
3-23
3-23