Fresenius Medical Care 4008 E Technical Manual page 250

Hemodialysis machine
Table of Contents

Advertisement

Part 3.1: Programmable mixing ratio (dilution)
Caution
The operator or technician is informed about his duty of care to enter the com-
ponent parameters and settings correctly or to check them and to set the CD
alarm window to the expected conductivity value.
When using the programmable mixing ratio (dilution), make sure to use the
right concentrate.
Using a bibag
not allowed.
Only enter authorized, programmable mixing ratios (dilutions).
SETUP DILUTION
Confirm
key
VARIABLE SETTING
®
in combination with the programmable mixing ratio (dilution) is
Confirm
B-Comp[part] : 0.800
key
Set the desired value
(0.800 – 2.500)
by pressing the
Override
key
ACKNOWLEDGED
After approx. 3 sec
RO[part] : 30.000
Set the desired value
(30.000 – 45.000)
by pressing the
Override
key
ACKNOWLEDGED
After approx. 3 sec
HCO3–[mmol]: 25
Set the desired value
(25 – 45)
by pressing the
Override
key
ACKNOWLEDGED
After approx. 3 sec
Na(B)[mmol]: 25
Set the desired value
(25 – 80)
by pressing the
Override
key
DATA STORED
After approx. 3 sec
Fresenius Medical Care 4008 4/08.03 (TM)
Select
key
keys.
Select
key
keys.
Select
key
keys.
Select
key
keys.
6-19