Fresenius Medical Care 4008 E Technical Manual page 277

Hemodialysis machine
Table of Contents

Advertisement

Part 7: Set dialysis parameters
SET DIAL PARAMETERS
Confirm
SET Flow Parameter
SET Temp. Parameter
SET Na/Bic Parameter
back to menu ?
Part 7.1: Set dialysate flow
SET Flow Parameter
Confirm
key
Flow[ml/min]: 500
Set the desired value
(300, 500, 800)
by pressing the +/– keys.
OR
Value set in Dial
Tone Mute
key
DATA STORED
After approx. 3 sec
Part 7.2: Set dialysate temperature
SET Temp. Parameter
Confirm
key
Temp.[°C]: 37,0
Set the desired value
(35.0 – 39.0)
by pressing the +/– keys.
OR
Value set in Dial
Tone Mute
key
DATA STORED
After approx. 3 sec
6-46
Fresenius Medical Care 4008 4/08.03 (TM)
key
Confirm
see Part 7.1
key
Confirm
see Part 7.2
key
Confirm
see Part 7.3
key
Confirm
key
Esc
key
Esc
key

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4008 b4008 h4008 s

Table of Contents