Fresenius Medical Care 4008 E Technical Manual page 177

Hemodialysis machine
Table of Contents

Advertisement

Part 6: Adjust UF pump volume
ADJ. UF-PUMP VOLUME
Confirm
optional
UF–Pump 1 (Vers. I)
UF-Pump 2
back to menu ?
Note:
Check the volume.
If necessary, readjust the UF pump
and repeat the procedure.
(See 3.3.5)
4-12
Fresenius Medical Care 4008 4/08.03 (TM)
key
Confirm
key
optional
Confirm
key
Confirm
key
UF I/O
key
Note
Graduated cylinder accuracy: ±0.5 %.
pulse-amount = 60
Enter the number of strokes
by pressing the
(+/–) keys.
Override
(Tone Mute)
key
ACKNOWLEDGED
approx. 3 sec
press uf key
Remove the line from the
UF pump, close the T-piece.
Hang the line into a
graduated cylinder.
Switch on the UF pump by
pressing the UF I/O key.
uf pulses left = 60
The remaining UF strokes are
indicated on the display.
The "UF Goal" display
indicates the number of
preselected strokes.
Selec
(Esc)
key
Select
(Esc)
key