Fresenius Medical Care 4008 4/08.03 (TM)
After the ONLINE system pump was switched off in the test, the actual
rate of the module is > 0 ml/min.
– Pump stop (output stage) defective
– Synchro-transmitter electronics defective
After activation of the substituate pump the monitoring unit (Hall sen-
sor) of the pump rotor detects incorrect rotation of the rotor (desired
rate 300 ml/min).
– Pump control defective (outside the tolerance of ±25%)
– Hall sensor /electronics defective
After activation of the substituate pump the system pump failed to
comply with the test rate of 300 ml/min ±9ml/min.
– Pump control defective
Port 1 open during ONL valve test sequence.
– Close port 1 (substituate port).
Port 2 open during ONL valve test sequence.
– Close port 2 (rinse port).
Acknowledgement of ONL1 (V193) differs from the desired state of the
valve.
– Valve control in the ONLINE Sys module defective
– Valve acknowledgement in the ONLINE Sys module defective
Acknowledgement of ONL2 (V192) differs from the desired state of the
valve.
– Valve control in the ONLINE Sys module defective
– Valve acknowledgement in the ONLINE Sys module defective
Acknowledgement of ONL3 (V191) differs from the desired state of the
valve.
– Valve control in the ONLINE Sys module defective
– Valve acknowledgement in the ONLINE Sys module defective
Leakage test ONL3 (V191) failed to be passed.
The permitted pressure drop of ∆P < –10 mmHg has been exceeded or
the test pressure is P ≤ 710 mmHg.
– Valve ONL3 (V191) in the ONLINE Sys module leaking
– Leaking system / tubing connections
– Port 1 or 2 in the ONLINE Sys module leaking
Leakage test ONL2 (V192) failed to be passed.
The permitted pressure drop of ∆P < –10 mmHg has been exceeded or
the test pressure is P ≤ 710 mmHg.
– Valve ONL2 (V192) in the ONLINE Sys module leaking
– Leaky system /tubing connections
– Port 1 in the ONLINE Sys module leaking
Leakage test ONL1 (V193) failed to be passed.
The permitted pressure drop of ∆P < –10 mmHg has been exceeded or
the test pressure is P ≤ 710mmHg.
– Valve ONL1 (V193) in the ONLINE Sys module leaking
– Leaky system /tubing connections