Fresenius Medical Care 4008 4/08.03 (TM)
Rinse phase has been aborted.
Valve V189 open (according to electronic acknowledgement), or failure
to perform 34 balancing chamber switchings within 240s.
– Valve control failed
– Failure to detect current rise pulse
– Conductivity not within the scale range
Possible cause: Concentrate and/or bicarbonate level sensor do not
recognize CD, although present.
See error message between F01 and F02 ONLINE plus /DIASAFE plus
See error message between F01 and F02 ONLINE plus /DIASAFE plus
See error message between F03 and F04 ONLINE plus /DIASAFE plus
See error message between F03 and F04 ONLINE plus /DIASAFE plus
See error message between F03 and F04 ONLINE plus /DIASAFE plus
System error