F33 ONLINE plus
F34 ONLINE plus
F34 DIASAFE plus
F41 ONLINE plus
F41 DIASAFE plus
F42 ONLINE plus
F42 DIASAFE plus
F43 ONLINE plus
F43 DIASAFE plus
F44 ONLINE plus
F44 DIASAFE plus
F95 ONLINE plus
F95 DIASAFE plus
F95 HPU
1-52
Fresenius Medical Care 4008 4/08.03 (TM)
Rinse phase has been aborted.
Valve V189 open (according to electronic acknowledgement), or failure
to perform 34 balancing chamber switchings within 240s.
– Valve control failed
– Failure to detect current rise pulse
– Conductivity not within the scale range
Possible cause: Concentrate and/or bicarbonate level sensor do not
recognize CD, although present.
See error message between F01 and F02 ONLINE plus /DIASAFE plus
See error message between F01 and F02 ONLINE plus /DIASAFE plus
See error message between F03 and F04 ONLINE plus /DIASAFE plus
See error message between F03 and F04 ONLINE plus /DIASAFE plus
See error message between F03 and F04 ONLINE plus /DIASAFE plus
System error