Checklist - Technical Safety Checks And Maintenance - Fresenius Medical Care 4008 E Technical Manual

Hemodialysis machine
Table of Contents

Advertisement

2.1.3
Checklist – Technical safety checks and maintenance
Customer/Customer no.:
Serial no.:
Machine type:
4008
With option: SN
BPM
1
Visual checks
TSC 1.1
Fuses accessible from the outside comply with the indicated values................................
TSC 1.2
Labels and identifications present and legible ...................................................................
TSC 1.3
Mechanical conditions permits further safe use ................................................................
TSC 1.4
No damage or contaminations detectable .........................................................................
TSC 1.5
Power cord not damaged ...................................................................................................
1.6
Preventive maintenance procedures
MA
1.6.1
Sealing in the suction tubes changed and lubricated, rivet replaced ................................
MA
1.6.2
Rubber in rinse chambers checked for proper function .....................................................
MA
1.6.3
Suction tube filters replaced ..............................................................................................
MA
1.6.4
Rinse chamber screws tight ...............................................................................................
MA
1.6.5
Check valve replaced ........................................................................................................
MA
1.6.6
Pre-UF pump filter, filter downstream of MV43, filter between rinse chambers,
and on MV99, MV100, CDS and disinfectant port replaced ..............................................
MA
1.6.7
Dialysate filter replaced or sieve changed .........................................................................
MA
1.6.8
O-rings in dialyzer couplings replaced ...............................................................................
MA
1.6.9
Sampling valve functions properly .....................................................................................
MA
1.6.10 Fan filter replaced ...............................................................................................................
MA
1.6.11 Ribbon belt and line segment in air separation pump changed .........................................
MA
1.6.12 MV84, replaced after 2 years. (Only if Puristeril is used.) .................................................
MA
1.6.13 Connecting piece or equilibration chamber replaced.
(Only if ONLINE™ plus or DIASAFE
MA
1.6.14 Filter 210 replaced (if present) ...........................................................................................
MA
1.6.15 Filter of the disinfectant suction tube replaced ..................................................................
MA
1.6.16 No dirty or worn tubings .....................................................................................................
2
General checks
TSC 2.1
Power failure alarm – continous sound – display: Emergency operation ..........................
TSC 2.2
Air separation by air separation pump activated;
display if more air must be separated and OD senses opaque fluid: fill program .............
TSC 2.3
V91, V99, V100 function properly and do not leak
(Check will not be applicable if CDS or hydraulics test is activated) .................................
3
Check of the hydraulics
MA
3.1
Water inlet pressure (reduced) 0.9 bar to 1.4 bar ................... Measured value:________
MA
3.2
Loading pressure 1.25 bar ±0.05 bar...................................... Measured value:________
MA
3.3
Negative degassing pump pressure 0.81 to 0.85 bar ............. Measured value:________
MA
3.4
Balancing chamber relief pressure at 800 ml/min
2.0 bar to 2.1 bar .................................................................... Measured value:________
Inventory no.:
4008 B
4008 H
bibag
®
4008 HDF
BVM
DIASAFE
4008 S
ON-LINE-HDF
DIASAFE
plus
plus
plus
plus
plus
®
®
plus option is not used.) ..............................................
Fresenius Medical Care 4008 4/09.03 (TM)
Service report no.:
Operating hours:
ONLINE plus
plus ™
plus
plus
plus
BTM
OCM
2-15