Fresenius Medical Care 4008 E Technical Manual page 46

Hemodialysis machine
Table of Contents

Advertisement

UF function test
Test description:
CPU1 activates the UF pump at a defined rate.
CPU2 checks the UF pump.
CPU2 blocks the control line of the UF pump and checks whether the UF pump stops.
Check of the UF counter.
The following is additionally applicable with built-in 4008 HDF option:
CPU1 activates the UF pump 2 at a defined rate.
CPU2 checks the hydraulic and the electric function of the UF pump 2.
CPU2 blocks the control line of the UF pump 2 and checks whether it stops.
Check of the UF2 counter.
Illustration: