T1 Test Flow Diagram, Parallel Program Steps - Fresenius Medical Care 4008 E Technical Manual

Hemodialysis machine
Table of Contents

Advertisement

1.1.2

T1 test flow diagram, parallel program steps

MODULE
START T1 TEST
TEST
OPT. DETECTOR
TEST OK
?
no
STORAGE
ERROR NUMBER
POS. PRESSURE
HOLDING TEST
ERROR NUMBER
CONDUCTIVITY
ERROR NUMBER
DIASAFE/HDF FILTER
ERROR NUMBER
T1
TEST
BLOOD SYSTEM
yes
yes
TEST OK
?
1
no
STORAGE
ERROR NUMBER
ERROR NUMBER
TEST
VENOUS PRESSURE
TEMPERATURE
yes
TEST OK
?
no
STORAGE
ERROR NUMBER
ERROR NUMBER
TEST
NEG. PRESSURE
AIR DETECTOR
HOLDING TEST
yes
TEST OK
?
no
1
STORAGE
ERROR NUMBER
ERROR NUMBER
no
yes
TEST OK
?
no
STORAGE
TEST
TEST
UF-FUNCTION
yes
TEST OK
TEST OK
?
?
no
no
STORAGE
STORAGE
ERROR NUMBER
TEST
yes
TEST OK
?
no
1
STORAGE
TEST
TEST
BYPASS
DISPLAY
yes
TEST OK
TEST OK
?
?
no
no
STORAGE
STORAGE
ERROR NUMBER
TEST
yes
TEST OK
?
no
STORAGE
TEST
BLOOD LEAK DETECTOR
yes
TEST OK
TEST OK
?
?
no
no
STORAGE
STORAGE
ERROR NUMBER
yes
Conductivity?
ERROR NUMBER
yes
ERROR NUMBER
ERROR NUMBER
Fresenius Medical Care 4008 4/08.03 (TM)
TEST
ACCUMULATOR
yes
yes
TEST OK
?
1
no
1
STORAGE
ERROR NUMBER
yes
1
TEST
CONDUCTIVITY
yes
TEST OK
?
no
STORAGE
POS. PRESSURE
HOLDING TEST
yes
TEST OK
?
no
STORAGE
TEST
TEST
UF-FUNCTION
DIASAFE/HDF FILTER
yes
TEST OK
TEST OK
?
?
no
1
STORAGE
STORAGE
ERROR NUMBER
TEST
ARTERIAL PRESSURE
yes
TEST OK
?
no
1
STORAGE
ERROR NUMBER
yes
no
1
1-5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4008 b4008 h4008 s

Table of Contents