Fresenius Medical Care 4008 E Technical Manual page 241

Hemodialysis machine
Table of Contents

Advertisement

Part 1: Set alarm and warning time
SET ALARM/WARN TIME
Confirm
Set ART-AL DELAYTIME
Set VEN-AL DELAYTIME
Set FLOW-OFF W-TIME
Set UF-WARNING-TIME
Set MUTE-TIME
back to menu ?
Part 1.1: Set the delay time of the arterial alarm
Set ART-AL DELAYTIME
Confirm
key
Art Al Delay = 5s
Set the desired delay time
(0 to 5 sec) by pressing
the
keys.
Override
key
DATA STORED
After approx. 3 sec
Part 1.2: Set the delay time of the venous alarm
Set VEN-AL DELAYTIME
Confirm
key
Ven Al Delay = 5s
Set the desired delay time
(0 to 5 sec) by pressing
the
keys.
Override
key
DATA STORED
After approx. 3 sec
6-10
Fresenius Medical Care 4008 4/08.03 (TM)
key
Confirm
see Part 1.1
key
Confirm
see Part 1.2
key
Confirm
see Part 1.3
key
Confirm
see Part 1.4
key
Confirm
see Part 1.5
key
Confirm
key
Select
key
Select
key