Part 7: Set dialysis parameters
SET DIAL PARAMETERS
Confirm
key
SET Flow Parameter
SET Temp. Parameter
SET Na/Bic Parameter
back to menu ?
Part 7.1: Set dialysate flow
SET Flow Parameter
Confirm
key
Flow[ml/min]: 500
Set the desired value
(300, 500, 800)
by pressing the
keys.
OR
Value set in Dial.
Override
key
DATA STORED
After approx. 3 sec
Part 7.2: Set dialysate temperature
SET Temp. Parameter
Confirm
key
Temp.[°C]: 37,0
Set the desired value
(35.0 – 39.0)
by pressing the
keys.
OR
Value set in Dial.
Override
key
DATA STORED
After approx. 3 sec
Confirm
see Part 7.1
key
Confirm
see Part 7.2
key
Confirm
see Part 7.3
key
Confirm
key
Select
key
Select
key
Fresenius Medical Care 4008 4/08.03 (TM)