Fresenius Medical Care 4008 E Technical Manual page 222

Hemodialysis machine
Table of Contents

Advertisement

O: CPU1_STOP_EP
Stopping of the degassing pump
IC 4/16
O: CPU1_STOP_FP
Stopping of the flow pump
IC 4/15
O: CPU1_SET_UF1_ON
Activation of UF pump 1
IC 4/14
(0/1 jump = 1 stroke)
O: CPU1_SET_UF2_ON
Activation of UF pump 2
IC 4/13
(0/1 jump = 1 stroke)
O: CPU1_SET_EN_UF2
Not used
IC 4/12
O: CPU1_SET_FLOW_ON
Flow on
Dataword to Gal 23:
0000 0010 (active, V 32 open)
0000 0011 (inactive, V 31, 32 open)
O: CPU1_SET_FILL_PRG
Fill program
Dataword to Gal 23:
0000 1010 (V 32, 34 open)
O: CPU1_EMPTIING_PRG
Emptying program:
Dataword to Gal 23:
0001 0010 (V 32, 35 open)
O: CPU1_FILL_ONE_CHAM
Filling of a balancing chamber com-
partment
Dataword to Gal 23:
0100 0010 (V 32, 37 open)
O: CPU1_EMPTY_ONE_CHA
Emptying of a balancing chamber com-
partment
Dataword to Gal 23:
1100 0010 (V 32, 37, 38 open)
O: CPU1_CO:L:XXXXXXXX
Step number of concentrate pump
The dataword to IC 2 is indicated on
the alphanumeric display and can be
changed by pressing the UF Rate UP/
DOWN keys.
Prerequisite: reed contact of concen-
trate connector open
O: CPU1_BI_L:XXXXXXXX
Step number of bicarbonate pump
Prerequisite: reed contact of bicar-
bonate connector open
O: CPU1_ALARM_SOUND
Alarm tone
IC 5/18, 19 set to 1: active
O: CPU1_WARN_SOUND
Warning tone
IC 5/18 set to 1, 19 to 0: active
O: CPU1_INFO_SOUND
Infosound
IC 5/18 set to 0, 19 to 1: active
O: CPU1_CLK_OVERLAP
Changeover of dead time of balancing
chamber
IC 5/17 set to 0: 1 kHz; to 1: 2 Hz
Fresenius Medical Care 4008 4/08.03 (TM)
5-23