Fresenius Medical Care 4008 E Technical Manual page 215

Hemodialysis machine
Table of Contents

Advertisement

I: CPU2_LDA2
LD alarm, channel 2
IC 5/7
Preparation:
LD alarm-free and set CLAMP_CTRL
(CPU1: WR DIGIT. OUTP) to 1
Check: generate an LD alarm
I: CPU2_BPSB_VEN
Venous blood pump stop confirmation
IC 5/8
Actuation of the Start/Stop key on the
venous blood pump
I: CPU2_BPSB_ART
Arterial blood pump stop confirmation
IC 5/9
Actuation of the Start/Stop key on the
arterial blood pump
I: CPU2_V42
Not used
IC 6/2
I: CPU2_BPST_ART
Acknowledgement, special control of
the arterial blood pump
IC 6/3
Check by setting BPST_ART
(CPU1: WR DIGIT. OUTP) to 1/0
I: CPU2_BPUS_ART
Arterial blood pump revolution stop
IC 6/4
Alarm field of the arterial blood pump
on = 0
(cleared by pressing the Start/Stop key
on the blood pump)
5-16
Fresenius Medical Care 4008 4/08.03 (TM)
I: CPU2_BPST_VEN
Acknowledgement, special control of
the venous blood pump
IC 6/5
Check by setting BPST_VEN
(CPU1: WR DIGIT. OUTP) to 1/0
I: CPU2_BPUS_VEN
Venous blood pump revolution stop
IC 6/6
Preparation:
Set SNST (CPU1: WR DIGIT. OUTP)
to 1 and wait for the alarm field of the
venous blood pump to turn on
Check: clear the alarm field by press-
ing the Start/Stop key on the venous
blood pump
I: CPU2_ADKS
Single needle blood pump recognition
connected
IC 6/7
Connect/remove the blood pump (with
the machine off only)
I: CPU2_LEVEL_UP
Raise LD level
IC 6/8, press the "Raise level" key
I: CPU2_LEVEL_DOWN
Lower LD level
IC 6/9
Preparation:
LD alarm-free, set CLAMP_CTRL
(CPU1: WR DIGIT. OUTP) to 1
Check: press the "Lower level" key
I: CPU2_RINSE
Concentrate reed contact
IC 7/2
Actuate rinse chamber/concentrate
reed contact