Part 5.3: Calibrate ONLINE-HDF pump
CALIB. HDF–PUMP–RATE
Confirm
calib. HDF–PUMP–RATE
HDF-Pump-Rate Check
back to menu ?
Note: This function is possible only if ONLINE-HDF has been activated by means of the DIP switch.
4-10
Fresenius Medical Care 4008 4/08.03 (TM)
key
Confirm
key
Confirm
key
Confirm
key
(B)P–Rate=200 ml/min (max)
Adjust various delivery
rates on the the HDF pump.
The values on display
and the HDF pump must be
identical.
Select
(Esc)
key
(B)P–Rate=400 ml/min
Enter the value of the
HDF pump display by
pressing the
(+/–) keys.
Override
(Tone Mute)
key
DATA STORED
approx. 3 sec
Select
(Esc)
key