Part 7: calibrate degassing pressure
CAL. DEGAS. PRESSURE
Confirm
key
adjust degas.-press.
Connect a pressure gauge for
the degassing pressure.
Set the degassing pressure
to –0.81 bar (to –0.85 bar) by
pressing the
(+/–) keys.
Override
(Tone Mute)
key
DATA STORED
approx. 3 sec
At this point, the following messages may appear:
– fill program active
– set flow on
See also 3.3.2 Degassing pump pressure
Select
(Esc)
key
Fresenius Medical Care 4008 4/08.03 (TM)
4-13