Without display messages: adjusting the blood pump stop alarm
(blood pump or HDF pump)
Set the hex switch in the module (P.C.B. LP 624, pos. 1) to position B.
Should the error message E02 appear on the blood pump display, clear the message by pressing
the Start/Stop key.
Change the preset value
(15/30 sec) by pressing
the
keys on the
blood pump module.
Acknowledge by
pressing the Start/Stop
key on the module.
The stored value appears
after 2 seconds.
Return the hex
switch to position 0.
Automatic reset
on the module
4-5
Fresenius Medical Care 4008 4/08.03 (TM)