Fresenius Medical Care 4008 E Technical Manual page 88

Hemodialysis machine
Table of Contents

Advertisement

Error codes:
E01
Hardware error, gate array defective
E02
Hardware error, reset by spike or test alarm
E03
Checksum error, data loss
E04
First start-up
E05
Incorrect hex switch position
E06
Missing or incorrect data for the variable syringe
E07
Selection of wrong syringe
E11 to E13 Step error
E12
Overdelivery during fast return
E14 to E15 Error in direction of rotation (software not equal to hardware!)
E16
Software error
E19
Optical sensor error (stop of syringe holder or optical sensor defective)
E20
Error in direction of rotation
E33
Step error (impermissible range)
E37
Slotted disc error
E40
Division error (division by zero)
E41 to E42 Error in direction of rotation (fast return)
E43 to E44 Error in direction of rotation (slow return)
E45 to E46 Error in direction of rotation (fast advance)
E47 to E48 Error in direction of rotation (slow advance)
E49
Step error (underdelivery during slow advance)
E50
Step error (underdelivery during slow return)
E51
Step error (overdelivery during fast advance)
E55
Error in step counting (optical sensor defective or mechanics too sluggish; no pulses
from the slotted disc)
E56
Error in step counting (more than 8 pulses during transition of the slotted disc; the
slotted disc is oscillating)
E90
Display error
Fresenius Medical Care 4008 4/08.03 (TM)
1-81