I: CPU1_V24
Valve 24
IC 13/5
Check by opening/closing the valve
I: CPU1_UF_P1
Acknowledgement of UF pump 1
IC 13/6
Check possible only in the combination
menu
I: CPU1_LDA2
LD alarm, channel 2
IC 13/7
Preparation:
LD alarm-free and set CLAMP_CTRL
(CPU1: WR DIGIT. OUTP) to 1
Check: initiate an LD alarm
I: CPU1_SUB_W_P
Acknowledgement of UF pump 2
IC 13/8
I: CPU1_LC11
IC 13/9
I: CPU1_REED_BIC
Bicarbonate reed contact
IC 14/2
Actuate rinse chamber/bicarbonate
reed contact
I: CPU1_BIBAG
Microswitch 137 / connector
IC 14/3
Check by connecting/removing the
BIBAG
I: CPU1_REED_RINSE
Concentrate reed contact
IC 14/4
Actuate rinse chamber/concentrate
reed contact
I: CPU1_BIBAG_C
Microswitch 138 / connector
IC 14/5
Check by connecting/removing
the cap
I: CPU1_PSW_V102
Concentrate pressure switch
IC 14/6
Check by increasing/decreasing the
pressure at the pressure switch
I: CPU1_PSW_V104
Bicarbonate pressure switch
IC 14/7
Check by increasing/decreasing the
pressure at the pressure switch
I: CPU1_PWR_OFF
Power off
IC 14/8
I: CPU1_HEP_ON
Heparin pump on
IC 14/9
Switch heparin pump on/off
I: CPU1_LA32
IC 15/2
Fresenius Medical Care 4008 4/08.03 (TM)
5-11