Heparin Pump Parameters - Fresenius Medical Care 5008 Operating Instructions Manual

Hemodialysis system
Hide thumbs Also See for 5008:
Table of Contents

Advertisement

Chapter 5: Preparation
5.4.11 Checking/Setting the Heparin Pump Parameters
5.4.12 Checking/Setting the Single-Needle Parameters
5-60
The following must be observed:
– The treatment may also be started with the UF profile only.
– It is only possible to select matching profile groups.
(E.g. if Na profile 1 has been selected, only UF profile 1 is available.)
– After having started the profiles, it is no longer possible to alter the
UF time and UF rate parameters.
– ISO UF parameters and Na parameters may be selected only before
starting or after ending a profile.
Set the desired parameters.
Touch the OK button to confirm the entered values.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
Caution
Administer the heparin dose according to the physician's instructions.
In the HEPARIN menu
Check the heparin pump parameters.
Set the desired parameters. Touch the OK button to confirm the values
entered.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
In the SINGLE-NEEDLE menu
Check the Single-Needle parameters.
Set the desired parameters. Touch the OK button to confirm the values
entered.
Visually check the confirmed values.
Touch the PREPARATION menu button to return to the
PREPARATION SCREEN.
Fresenius Medical Care 5008 OP 5/09.06