Konica Minolta SONIMAGE HS2 Operation Manual page 78

Ultrasound system
Table of Contents

Advertisement

2 .2 Power Supply Cord Connection and Battery Capacity
z Battery status indicator
The system monitors the status of the battery's re-
maining capacity / AC adapter connection, and informs
you of changes, if any, in each status by varying the
on-screen status indicator icon and the status indicator
LED on its mini console according to such changes .
AC adapter connection
• Connected
Battery's remaining
capacity
100 %
99 % to 75 %
74 % to 50 %
49 % to 25 %
24 % to 0 %
Disconnected
Status is unknown
Battery error
• Disconnected
Battery's remaining
capacity
100 %
99 % to 75 %
74 % to 50 %
49 % to 25 %
24 % to 0 %
Disconnected
Status is unknown
Battery error
Status indicator icon
(green)
(green)
(green)
(yellow)
(red)
Status indicator icon
(green)
(green)
(green)
(yellow)
(red)
• Unknown
Battery's remaining
capacity
• When the remaining battery level falls to 15 % or
less, a message appears, prompting you to connect
the AC adapter .
• When the remaining battery level falls to 5 % or less,
the forced shutdown message appears and then the
system shuts down .
IMPORTANT
••••••••••••••••••••••••••••••••• •• • •
• When an error occurs, a message is displayed .
Follow the message instructions .
• When no battery has been mounted, or when you
remove the battery during start-up or operation, a
message appears, prompting you to make battery
connections .
• When setting the battery low notification to off,
check the status indicator icon for battery level
while using this system . The system may shut
down during operation if you are not aware of low
battery status .
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
••••••••••••••••••••••••••••••••• •• • •
HINT
• The system can also be set to notify you of low
battery using sound instead of a message . Refer to
the item "Low battery notification" under "System
Screen" in the "Operation Manual: Applications" .
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
z The storage temperature range and period
for batteries
Store batteries within the temperature range specified
below:
- 20 °C to 60 °C (30 days)
- 20 °C to 40 °C (90 days)
- 20 °C to 20 °C (1 year)
IMPORTANT
••••••••••••••••••••••••••••••••• •• • •
• The battery has a limited lifetime . If the operating
time becomes shorter, replace the battery with a
new one .
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
78