Color Flow-mode parameter list
Item
Sets the transmission voltage .
[Tx Level]
• Displayed in all modes .
Velocity setting for Color Flow-mode
[C Scale]
• The parameter range depends on the transducer used .
Steer angle setting for Color Flow-mode
[C Steer]
• Enabled when a linear transducer has been connected to the system .
[C Gain]
Gain setting for Color Flow-mode
[C Filter]
Setting for decreasing the low-velocity noise of Color Flow-mode
Frequency setting for Color Flow-mode
[C Frequency]
• The actual frequency values differ from one transducer to another.
[Temporal]
Repeated transmission count setting for Color Flow-mode
Acoustic line density setting for Color Flow-mode
[C Density]
• IL can be selected if a linear transducer or sector transducer is connected .
[Motion Sup .]
Motion artifact reduction setting for Color Flow-mode
[Spatial Avg .]
Spatial direction smoothing setting for Color Flow-mode
[C Time Avg .]
Time direction smoothing setting for Color Flow-mode
Display priority setting for Color Flow-mode
[C Priority]
• A larger value represents a higher priority, being more effective in overriding B-mode.
[C Invert]
Color invert setting for Color Flow-mode
[C Baseline]
Baseline setting for Color Flow-mode
Map setting for Color Flow-mode
[Color Map]
• When [Improvement of Noise Reduction Performance in CFM/PDI/SCF Mode] on the settings
screen is on, you cannot choose [VV1] to [VV4] .
Changes the B-mode parameters in order to improve Color Flow-mode frame rate .
[FR priority]
• When [FR Priority] is set to anything other than [Off], the setting values of THI and acoustic line
density may be limited .
[C SI-Filter]
Image processing setting for Color Flow-mode
Sets a process to improve resolution for CFM/PDI/SCF mode .
[C iXRET]
• The parameter range depends on the transducer used .
The following adjustments are done in the relevant modes .
• Auto C ROI adjustment in Color Flow-mode
[Simple Adjust]
• Auto Doppler gate adjustment in PW Doppler-mode
• Auto flow velocity adjustment in PW Doppler-mode/CW Doppler-mode/PW-TDI-mode
Functions
145