4 .4 Basic Scanning Mode
2
When [Ana-M] is set to [On], the freely movable Ana-M cursor and the rotation center cursor
appear .
Ana-M cursor
• The Ana-M cursor can be rotated centered on the rotation center cursor by the Angle operation (the item in the M-
mode menu or F2 Rotate) .
• The Ana-M cursor is displayed as a straight line from one side to another side of the B image . Among the intersec-
tions of the line (cursor) with the sides, the one corresponding to the end position of the M image is displayed as
an arrow . Among the intersections, the one always having a deeper display depth for the B image is the end posi-
tion side . (When the cursor's rotation angle is 180 degrees and horizontal to the B image, the start and end posi-
tions depend on the setting of the B-mode parameter, [LR Flip] . When [LR Flip] is set to [Left], the left side is the
start position and the right side is the end position .)
• The rotation center cursor is displayed at the center of the Ana-M cursor . It can be moved with trackball or touch
operation .
• When the Anatomical M-mode is activated in a frozen state, the Anatomical M image is displayed from Cine mem-
ory in B-mode .
Rotation center cursor
<Anatomical M image>