z Essential performance of this system
• Free from noise on a waveform or artefacts or distortion in an image or error of a displayed numerical value which cannot be at-
tributed to a physiological effect and which may alter the diagnosis.
• Free from the display of incorrect numerical values associated with the diagnosis to be performed .
• Free from the display of incorrect safety-related indications .
• Free from the production of unintended or excessive ultrasound output .
• Free from the production of unintended or excessive TRANSDUCER ASSEMBLY surface temperature .
Notice1: "artifact" means "noise" that generates acoustically .
Notice2: "incorrect" in the sense that the displayed value differs from what is calculated (having been altered during
data transfer), or the calculation itself is not correct .
z List of transducers for conformance standard of EMC
Transducer name
C5-2
L18-4
S4-2
L14-4
EC9-3
L11-3
HL18-4
MC10-3
WL13-3
S4-2A
z List of cables for conformance standard of EMC
Cable name
Power supply cord
(Hospital Grade)
USB cable
LAN cable
DVI cable
ECG cable
11 .6 Electromagnetic Emissions and Electromagnetic Immunity
Shielded/Unshielded
Shielded
Shielded
Shielded
Shielded
Shielded
Shielded
Shielded
Shielded
Shielded
Shielded
Shielded/Unshielded
Unshielded
Shielded
Shielded
Shielded
Shielded
303
Length
< 3 m
< 3 m
< 3 m
< 3 m
< 3 m
< 3 m
< 3 m
< 3 m
< 3 m
< 3 m
Ferrite core
Not exist
Exist
Not exist
Exist
Exist
Length
≤ 3.5 m
≤ 2.0 m
≤ 10 m
≤ 2.0 m
≤ 3.5 m