Konica Minolta SONIMAGE HS2 Operation Manual page 115

Ultrasound system
Table of Contents

Advertisement

Restore
This function is used to repair or restore all or part of data .
There are two Restore methods .
• Restoring the settings for the system
• Restoring patient / study data
z Restoring the settings for the system
1
Press the [Restore Setup] button on the [Backup / Restore] screen .
• The screen changes to the [Restore Setup] screen .
• Connect an external storage medium .
2
Select a backup destination from the [Device] pull-down menu .
• When [Temporary Backup] of [Device] has been selected, data is restored from Internal Storage .
3
Select data to be backed up from the [Backup Data] pull-down menu .
4
Check a category box to be restored .
5
Press the [Restore] button .
• When restoring data after having deleted the system's data, check the [Delete system setup data before re-
store .] box before pressing the [Restore] button .
• When a restoring operation has been completed, a message screen appears . When you press the [Restart]
button, the system restarts .
• To return to the [Backup / Restore] screen, press the [Back] button .
Step 2
Step 3
115
3 .7 Data Backup
Step 4
Step 5