• When recording diagnostic images, confirm that
the displayed patient ID and patient are matched .
Recording diagnostic images with an incorrect pa-
tient ID may result in incorrect diagnosis . If the error
message is displayed, the power switches of the
peripheral units and the cables between the system
and the peripheral units must be checked . Contact
your service representative .
• Clips (AVI format) recorded in the internal storage
are always irreversibly compressed . Irreversible
compression is effective for decreasing the volume
of an image, but the quality of the image deterio-
rates in comparison with when the image is not
compressed . Use irreversible compression only
when you have assessed that it does not affect di-
agnosis .
• It is also possible to perform a measurement on a
saved image on which another measurement was
made during the examination and to print out the
result . To avoid being confused and misled by the
printed image as it will be viewed later, take ap-
propriate measures, such as keeping a record of or
making a mark on printed images, for ease of iden-
tification . In the case of black-and-white printing,
a measurement or the result of the measurement,
which was already included in a saved image, may
be confused with another measurement, which was
newly made on the saved image, or the result of
another measurement .
CAUTION
Wireless LAN
• Wireless LAN adapter of the system is for indoor
use only . Please remove the wireless LAN adapter
from the system in advance, if the system may be
in the outdoor locations .
• Do not use the wireless LAN adapter in the environ-
ment occurring of magnetic field, electrostatics or
electromagnetic interference, etc . If wireless LAN
connection is cut due to those interferences, use
wired LAN connection .
• When similar wireless LAN adapter exists around
the system, it may occur mutual interference .
Therefore, install the system after inspection of
other wireless LAN to avoid interference .
• For Warnings and Cautions regarding the wireless
LAN adapter, refer to the operation manual of the
wireless LAN adapter .
• Where the strength of the wireless signal is weak,
please move to a place of strong signal or use
wired LAN connection .
• It is recommended to connect the wireless LAN
adapter to the USB connector on the lateral side
of the system . The communication rate may be
reduced depending on the positional relationship
between the wireless LAN adapter and the access
point .
Regarding data transfer in weak wireless signal
connection
• When transferring several files, transfer of the cur-
rent file cannot be interrupted . Wait for the trans-
fer to be completed . After current transferring file
completes data transfer, data transfer can be inter-
rupted .
Regarding operations during start-up
• When the system is starting up, avoid touching the
LCD monitor until the start-up process is complete .
Otherwise, the sensitivity of the touch panel cannot
be adjusted accurately .
• Do not press the optional USB keyboard during
start-up . Otherwise, the system may malfunction .
34