Konica Minolta SONIMAGE HS2 Operation Manual page 153

Ultrasound system
Table of Contents

Advertisement

Item
Select the Display method of the guide display for checking the pressurization state
• [Off]: No display
[Navi Type]
• 1: Display the time variation of the strain
• 2: Display the repetition tempo score for pressurization and relaxation
• 3: Display the pressurization stability
[Navi Position]
Select the Display location of the guide display to check the pressurization state
Select the Color Map of the Left or Upper screen in the Single and Split-screen displays
[Color Map]
Stronger coloration in the upper color map indicates softer tissue . Stronger coloration in the lower
[Color Map L(U)]
map indicates harder tissue .
[Color Map R(D)]
Select the Color Map of the right or lower screen in the Split display
Select the color level of elastography image
[Color Offset]
• The smaller the value the stronger the blue color: the larger the value the stronger the red color .
Select the elastography image color range
[Color Range]
• The smaller the value the stronger the color contrast .
Select the frequency for Elastography-mode
[E Frequency]
• [eGen] : For general purpose
• [ePen] : For penetrated part
[E Invert]
Color display inversion
Select the scanning line density
[E Density]
• The scanning line density increases in the order [L] < [M] < [H] reducing the frame rate .
Select the pressurization speed
[Comp . Speed]
• The larger the value, the higher the frame rate . A higher frame rate is required to match a higher
pressurization speed .
Select the Time Avg . strength
[E Time Avg .]
• The larger the value the smoother the time variation of the elastography image .
Select the Spatial Avg . strength
[Spatial Avg .]
• The larger the value the smoother the spatial variation of the elastography image .
Select the strength for removing the noise frame
[Rejection]
• The larger the value the greater the noise reduction .
Select the strength for homogenizing uneven pressurization of the elastography image
[Equalization]
• The larger the value the greater the contrast is enhanced/unevenness of the elastic image is ho-
mogenized .
Select the blending ratio for the elastography and B-mode images
[Blend Ratio]
• The larger the value the greater the composition ratio of the elastography image .
Switches the stable pressurization interval display in Cine data
• Analyzes the most stable pressurization state in Cine data and displays it on the Cine bar in the
freeze state .
• If there is no continuous stable pressurization interval no stable interval is displayed .
[Stable Section]
• When a frame inside the stable interval is selected the following display changes .
- The Cine numbers are displayed in green .
- The green frame is displayed in pressurization state navigation .
• The approx . 1-second stable interval is displayed .
Functions
153