Freely specifying the relationship between display values and analog output values
PREPARATION
(equivalent to the former ZX-L-N monitor focus)
FOR
MEASUREMENT
• To specify any analog output value for a display value, assign the analog output range and
the minimum and maximum analog output values by selecting the analog output and then
FLOW OF
OPERATION
setting up scaling.
(If scaling is not set up, the measurement range is the same as the analog output range.)
BASIC
SETUP
The analog output range is assigned based on the post-scaling display value setting range
(between S1-AFT and S2-AFT).
MAIN
APPLICATIONS
Concerning the minimum and maximum analog output values, the analog output minimum
& SETTING
METHODS
value is output for the smaller of the post-scaling display values (S1-AFT/S2-AFT), and the
Height
analog output maximum value is output for the larger of these values.
Steps
and
• To only specify the analog output range, without changing display values
Warpage
Example: To set the analog output in the range of –5 V to 5 V
Double
Sheet
Detection
Thickness
(1) Select
(2) Specify the measurement range to use for the BEF and
AFT values, and then assign the analog output range
Positioning
based on the measured value range.
Eccentricity
• S1-BEF: –5 (mm)
and Surface
Deflection
• S1-AFT: –5 (mm) → Set the same value as S1-BEF
• S2-BEF: 5 (mm)
DETAILED
• S2-AFT: 5 (mm) → Set the same value as S2-BEF
SETTINGS
<Initial setting>
TROUBLE-
SHOOTING
Display value Analog output
SPECIFI-
CATIONS
–10 mm
10 mm
INDEX
• To specify the analog output range after changing display values
SETTING
(For details on scaling, see page 108.)
TRANSITION
CHARTS
110
Analog Output
Display
Description of Operation
Hold down the
Out
three seconds to switch to the
RUN mode.
when using the ZX2-LD50(L) at a distance of 45 mm
to 55 mm from the sensor:
as the analog output setting.
–5 V
5 V
button for
<Scaling setting>
Pre-scaling
Scaling point
display value
(BEF)
S1
–5 mm
S2
5 mm
Explanation of
Selection Menu
5 V
-5 mm
5 mm
-5 V
Post-scaling
display value
Analog output
(AFT)
–5 mm
–5 V
5 mm
5 V