P E R FO RMANCE TE STS
A. Connect the test instruments as shown in figure 2-25. Keep the leads from the
Proced u re
feedthrough term inations to chassis ground as short as possible.
B. Press the HP 3562A keys as fol lows:
PRESET
CAL
RA N G E
I N P U T
CO U P LE
F REQ
AVG
W I N DOW
U N ITS
C. Perform steps 1 through 4:
1 . Press the H P 3562A keys as fol lows:
START
SCALE
SPC L
MARKER
2. If the Ya marker read ing is less than or equal to - 1 31 dBVrms, check PASS on the
Perform ance Test Record for CHAN 1 .
2-74
R E S E T
S I N G L E
CAL
d BVrms
- 51
CHAN 1
AC
CHAN 2
AC
FREQ S PAN
START FREQ
STABLE
20
U N I FRM
P S PEC
UN ITS
Y A UTO
SCALE
MRKR .... .
PEAK
MO D E L 3562
kHz
Hz
1
E NT E R
2 0
VOLTS
RMS
VOLTS