3. Press the HP 3562A keys as fol lows:
B
SCALE
SPCL
MARKER
4. If the Yb marker read ing is less than or equal to - 1 31 dBVrms check PASS
on the Operation Verification Test Record for CHAN 2.
Table 2-3 (OV) Spurious Signals
Start
Frequency
20 Hz
1 kHz
90 kHz
D. For the rest of the start frequencies in tab le 2-3 perform steps 1 through 4:
1 . Press the HP 3562A keys as fol lows:
F REQ
A
START
SPCL
MARKER
2. If the Ya marker read ing is less than or eq ual to - 1 31 dBVrms check PASS
on the Operation Verification Test Record for CHAN 1 .
Y AUTO
SCA LE
MRKR -
PEAK
Frequency
Span
Specification
:;:; - 1 31 d BV
1 kHz
10 kHz
:;:; - 1 31 d BV
1 0 kHz
:;:; - 1 31 d BV
START FREQ
FREQ SPAN
MRKR -
P EAK
O P E RAT I ONAL V E R I FICAT ION
To start freq uency in
table
To freq uency span in
table
2-23