2. Press the HP 3562A keys as fol lows:
F REQ
RA N G E
START
SPCL
MARKER
3 . Record the Ya marker amplitude reading o n the Operation Verification Test Record
as the fi rst measurement for C HAN 1 .
4. Record the Yb marker ampl itude reading on the Operation Verification Test Record
as the fi rst measurement for C HAN 2.
FREQUENCY SYNTHESIZER
HP 3325A
8 8 8.
00000
88
00000
00000
00000
8
o
Figure 2-7 (CV) Common Mode Rejection Test Setup #2
C E NT E R FREQ
To range setting #1 i n table
MRKR
P EA K
HP 3562A
: : : I tIJ
8
8 8
8 ·
o
III
BNC/BNC
B120 1B40
O P E RATI ONAL V E RI F I CATION
To signal frequency
in table
@
� � �-.. -�
0 000 000
COMMON MODE
HP 03562-61620
BNC CABLE
B120-1B3B
'-TEE
1250-07B1
'-F TO
F BARREL
-OOBO
1250
2-1 7