HP 3562A Service Manual page 63

Dynamic signal analyzer
Hide thumbs Also See for 3562A:
Table of Contents

Advertisement

I N P U T
CO U P LE
U N ITS
A & B
COO RD
SCA LE
Table 2-9 Amplitude Linearity
AC Calibrator
Amplitude
Upper limit
1 0.00 Vrms
1 .000 Vrms
1 00.0 mVrms
1 03.2 mVrms
1 0.00 mVrms
1 1 .67 mVrms
3 .1 623 mVrms
4.71 7 mVrms
D. For each of the ampl itudes listed in table 2-9 perform steps 1 through 4.
1 .000 mVrms
2.51 7 mVrms
1 . Set the ac calibrator's ampl itude.
2 . Press the H P 3562A keys a s fol lows:
START
SPCL
MARKER
3. Record the Ya marker read ing on the Performance Test Record for the
measu red value CHAN 1 .
Record the Yb marker read ing on the Performance Test Record for the
measu red value CHAN 2.
G ROUND
CHAN 1
G RO U N D
C H A N 2
P SPEC
UN ITS
MAG
(U N)
Y AUTO SCA L E
Specification
BNC shell grounded
Lower limit
1 0.1 8 Vrms
9.827 Vrms
1 .01 9 Vrms
981 .4 mVrms
96.79 mVrms
8.329 mVrms
1 .608 mVrms
- 51 7.1 uVrms
MRKR
PEAK
P E R FORMANCE T ESTS
VOLTS
RMS
VOLTS
Specification
BNC center conductor grounded
Upper limit
Lower limit
1 0.59 Vrms
9.439 Vrms
1 .061 Vrms
942.6 mVrms
1 07.4 mVrms
92.91 mVrms
1 2.09 mVrms
7.941 mVrms
4.850 mVrms
1 .485 mVrms
2.559 mVrms
- 555.9 uVrms
2-41

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents