HP 3562A Service Manual page 515

Dynamic signal analyzer
Hide thumbs Also See for 3562A:
Table of Contents

Advertisement

S E RV I C E
D. Signature Analysis Tests
The fol l owi ng six signature analysis tests are des igned to test c i rc u it b l ocks of the d igital
fi lter assemb ly. Perform the fo l lowi n g steps to configure the 3561 A for any of the s ix tests :
Discon nect the power cable.
Put the board u nder test on an exte nder card . A l l j u m pe rs shou ld be i n the normal
(N) position.
Connect and configu re the signature analyzer as descri bed at the beg i n n i n g of the
test you wish to perform.
Move j u m pers as instru cted at the beg i n n i n g of the test.
Connect the power cab le and turn on power.
I n itiate the D FA PATT test as i n stru cted at the beg i n n i n g of the test. To put a test
in the loop mode, press the SPCL FCTN key, select S E RV I C T E ST, and p ress the LOO P
menu key to l ight u p O N .
1 .
Signature An alysis Test AS #1 System I nterface
a.
Put A5J 5 and ASH i n test pos ition . Put A6J 2 i n test pos iti on.
b .
Connect the signature analyzer a s fol lows:
c.
Put D FA PATT 1 i n Loop Mode
C heck the signatu res i n tabl e A5-2 .
d.
Signal
Polarity
C N D
START
+
-
STO P
C LOCK
+
&
Control
Connection
AS TP8
AS TP1 3
AS TP1 3
AS TPS ,
MO D E L 3562A
8-65