A28 FDD Replacement

Tools Required

TORX screwdriver, T10, T15, and T20

Removal procedure

Refer to Figure 6-9 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 176.
Step 2. Remove the front panel as described in "Front Panel Removal" on page 178.
Step 3. Disconnect the power supply cable from the FDD.
Step 4. Disconnect the flat cable from the FDD.
Step 5. Release the clamped cable from the clamp (item 1) of under the FDD.
Step 6. Remove four TORX T10 screws (item 2) fastening the FDD assembly.
Step 7. Slide the FDD assembly forward.
Step 8. Remove the four TORX T10 screws (item 3) fastening the FDD to the mount.
Replacement Procedure
Step 1. Fasten the angle to the new FDD, and then fasten it to the mount as shown in Figure 6-9.
Step 2. Reverse the order of the removal procedure.
Figure 6-9
A28 FDD Removal
Chapter 6
Replacement Procedure
A28 FDD Replacement