Pre-Lo Sampler Pll Bw Adjustment; Receiver Dc Offset Adjustment; Pm Offset Adjustment - Agilent Technologies E5052A Service Manual

Signal source
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Required equipment for reference frequency adjustment
Description
Signal Generator
N(m)-N(m) Cable

Pre-LO Sampler PLL BW Adjustment

The purpose of this procedure is to pre-adjust the LO sampler PLL BW of the A4/A5 local
synthesize modules.
Required equipment for the Pre-LO sampler PLL BW adjustment
Description
Signal Generator
N(m)-N(m) Cable

Receiver DC offset Adjustment

The purpose of this procedure is to adjust the DC offset of the Receiver section (A6 Phase
detector board and A3 ADC/DSP board).
Required equipment for the receiver DC offset adjustment
Description
50 ohm (N) Termination
50 ohm (BNC) Termination

PM Offset Adjustment

The purpose of this procedure is to adjust the PM offset of the A1 freq/power detector
module.
Chapter 3
Required Adjustment after Replacing Assembly
Recommended Model
Agilent E8257D option UNR,
UNT
Agilent N6314A (p/n
8120-8862)
Recommended Model
Agilent E8257D option UNR,
UNT
Agilent N6314A (p/n
8120-8862)
Recommended Model
Agilent 909C opt.012
Agilent 11593A
Adjustment
77