Figure 4-20
Opening Device Driver Property

To Check the Front Panel

Procedure
Randomly press the front panel keys and rotate the knob to verify that they work normally.
Step 1. Press
Step 2. Click
Step 3. Click
Figure 4-21.
Figure 4-21
Front Panel Test dialog box
Step 4. Randomly press the front panel keys. The key code along with the name of the pressed key
are displayed in the dialog box as shown in Figure 4-22. Turn the rotary knob clockwise or
counterclockwise. The dialog box indicates the direction of the turned knob and a count of
RPG output.
Chapter 4
key.
and, then,
Service Menu
Test Menu
in the test menu. This opens "Front Panel Test" dialog box as shown in
Front Panel
Function Specific Troubleshooting
in the softkeys.
Troubleshooting