Replacement Procedure; A1 Freq / Power Detector Module Replacement; Tools Required; Removal Procedure - Agilent Technologies E5052A Service Manual

Signal source
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Replacement Procedure

A1 Freq / Power Detector Module Replacement

A1 Freq / Power Detector Module Replacement

Tools Required

Removal Procedure

Refer to Figure 6-13 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 176
Step 2. Remove the front panel as described in "Front Panel Removal" on page 178.
Step 3. Disconnect the semi-rigid cables and RF cable(item 1).
Step 4. Remove the TORX T15 screw fastening the freq / power detector module bracket (item 2).
Step 5. Slide out the freq / power detector module (item 3).
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
Fasten the semi rigid cable using a torque wrench for replacement.
Figure 6-13
A1 Freq/Power Detector Module Removal
194
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Chapter 6