Adjustment
Required Adjustment after Replacing Assembly
Required equipment for DC control voltage adjustment
Description
Multimeter
BNC(f)-Banana Adapter
BNC(m)-BNC(m) Cable, 61 cm

DC Power Voltage Adjustment

The purpose of this procedure is to adjust the DC power voltage of the A2 DC source
board.
Required equipment for the DC power voltage adjustment
Description
Multimeter
BNC(f)-Banana Adapter
BNC(m)-BNC(m) Cable, 61 cm

DC Power Current Adjustment

The purpose of this procedure is to adjust the DC power current measurement of the A1
freq/power detector module.
Required equipment for the DC power current adjustment
Description
Multimeter
141 ohm Load
BNC(f)-Banana Adapter
BNC Tee Adapter
BNC(m)-BNC(m) Cable, 61 cm

Reference Frequency Adjustment

The purpose of this procedure is to adjust the reference frequency of the A4/A5 local
synthesize modules.
76
Recommended Model
Agilent 3458A
Agilent p/n 1251-2277
Agilent p/n 8120-1839
Recommended Model
Agilent 3458A
Agilent p/n 1251-2277
Agilent p/n 8120-1839
Recommended Model
Agilent 3458A
Agilent p/n 0698-3615 (47 Ω),
3ea
Agilent p/n 1251-2277
Agilent p/n 1250-0781
Agilent p/n 8120-1839