Replacement Procedure; Fan Replacement; Tools Required; Removal Procedure - Agilent Technologies E5052A Service Manual

Signal source
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Replacement Procedure

Fan Replacement

Fan Replacement

Tools Required

Removal Procedure

Refer to Figure 6-18 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 176.
Step 2. Remove the front panel as described in "Front Panel Removal" on page 178.
Step 3. Disconnect the fan cables (item 1) from the analog board.
Step 4. Remove the four white rivets (plastic)(item 2) fastening the fan.
Step 5. Remove the four black rivets (plastic)(item 3) fastening the fan using the fine stick.
Figure 6-18
Fan Removal
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
200
TORX screwdriver, T10,T15 and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
A fine stick such as Hex key, 1.5mm
Chapter 6