USB (USBTMC) Interface Card Replacement for E5052A Serial Prefix MY442
USB (USBTMC) Interface Card Replacement for E5052A
Serial Prefix MY442 and above

Tools Required

TORX screwdriver, T10, T15, and T20

Removal Procedure

Refer to Figure 6-24 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 176.
Step 2. Remove the three TORX T10 screws (item 1) fastening the angle.
Step 3. Remove two TORX T10 screws (item 2) fastening the PCI DSP Card/GPIB Card assembly
to the rear panel.
Step 4. Lift the USB (USBTMC) interface card (item 3) from the Digital Motherboard.

Replacement Procedure

Step 1. To replace the USB (USBTMC) interface card, reverse the order of the removal procedure.
NOTE
It is not necessary to replace the angle that comes with the replaceable part for the USB
interface. Replace only the board and use the existing angle. However, should it be
necessary the angle, make sure the gasket (0.012 m (2 pieces)) are fitted.
Chapter 6
Replacement Procedure
and above
209