A6 Phase Detector Board Replacement

Tools Required

TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)

Removal Procedure

Refer to Figure 6-17 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 176.
Step 2. Remove the front panel as described in "Front Panel Removal" on page 178.
Step 3. Disconnect the all semi-rigid cables (item 1).
Step 4. Disconnect the RF cable (item 2).
Step 5. Remove two TORX T10 screws (item 3).
Step 6. Slide out the Phase Detector board (item 4).
Figure 6-17
A6 Phase Detector Board Removal
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
Fasten the semi rigid cable using a torque wrench for replacement.
Chapter 6
Replacement Procedure
A6 Phase Detector Board Replacement
199